Administrator/Service run diagnostics
The vTanis unit supports a number of interactive diagnostic tests designed to check specific areas of
operation of the device. The tests are all designed to be executed locally so the unit’s behavior can
be observed. The tests are as follows:
Slot UI Test:
It tests each slot for eject button indication and sensing- requires a local operator to interact with the
unit.
Display Test:
It tests the LCD display and backlight operation- requires a local operator to step the display with the
OCP navigation button and observe the display through various states to confirm operation.
Fan Test:
It tests the operation of the vTanis cooling fan, steps the fan speed through four RPM ranges while
monitoring fan rpm.
Cartridge Access Test:
It tests the operation of each slot location by checking communication with a test cartridge.
Execution of this test will force the unit offline and eject all inserted cartridges (this is a data safety
feature to insure the test operator can use a scratch cartridge) from the unit- requires a local
operator to interact with the unit.
To start the diagnostic test sequence select each test by checking each test from the list that you
would like to perform and push the “Start Diagnostic” button. You may select to run only one test or
all four, upon completion of the selected test(s) a new diagnostic log status page will be created
capturing the current test status